Luminex 100 IS User Manual Version 2.3

x

MAP Technology

7 - 16

PN 89-00002-00-072 Rev. C

Verify that the system is
functioning properly by running
the system controls. If the
system controls pass, contact the
kit manufacturer.

Failed Curve Fit

The calculation could not occur
for the assay standards for data
interpretation.

Verify that you selected the
correct template.

Verify that the correct control
lot number and target values
were selected and entered
correctly.

Verify that the system is
functioning properly by running
the system controls. If the
system controls pass, contact the
kit manufacturer.

Sample High or Low

The results do not fall within the
highest and lowest standards.

Verify that you selected the
correct template.

The wrong control lot number
or target values are selected.

Verify that the correct control
lot number and target values
were selected and entered
correctly.

Verify that the system is
functioning properly by running
the system controls. If the
sample is noted as High, dilute
the sample following the kit
manufacturer’s suggestions. If
the sample is noted as Low,
vortex the sample to ensure
homogeneity.

Sample Timed Out

The samples are concentrated.

Verify that the samples are
properly mixed.

Not enough sample loaded into
the well.

Verify that you loaded the
correct sample volume to the
well.

Sample probe is clogged.

Remove and sonicate or flush
the sample probe to remove any
clogs.