Change lot -85 – Luminex 100 IS Version 2.3 User Manual

Page 127

Advertising
background image

x

MAP Technology

Using Luminex 100 IS 2.3 Software

PN 89-00002-00-072 Rev. C

5 - 85

1. Menu for multiple tests 2. Menu for single test

Figure 5-54. Invalidate and Validate Shortcut Menus

2. Click Recalc to recalculate the results if the Auto option was not

selected.

‹

To validate or invalidate a standard or control entry using the

Invalidate and Validate Buttons:

1. Select the desired standard or control name in the Standards or

Controls grid.

2. Click the appropriate buttons at the bottom of the Analysis

window: Invalidate Standard (F4), Validate Standard (F5),
Invalidate Control (F6), or Validate Control (F7).

3. The appropriate Standard or Control dialog box opens. To

invalidate or validate all tests click Yes. For only the single test
click No. When invalidating, the Name box turns red with an
asterisk proceeding it. When validating, the Name box reverts to
black text.

4. Click Recalc to recalculate the results if the Auto option was not

selected.

Change Lot

Use the Change Lot command to edit the lot that is applied to the
batch currently opened in data analysis.

‹

To change the lot to another available lot:

1. Click Change Lot (Alt + F8), located at the bottom of Analysis

window, to display the Choose Lot dialog box. See Figure 5-55.

2. The dialog box displays a list of available standard and control

lots that you can apply to a batch. Highlight the desired lot and
click OK to apply the selected lot to the batch opened in data

1

2

Advertising