Returning to the weighing mode – RADWAG PS 510.R2.CT Carat Balance User Manual

Page 135

Advertising
background image

- 135 -

In order to operate properly, you should set the threshold in grams. It is

the indicating point beyond which the function starts to register

maximal force apllied.

From now on the balance registers and holds every single weighment

which is above the threshold, and is higher than the result of the

previous peak hold. If the software detects a mass above the

threshold, the highest detected indication is held on the main display

and the pictogram <Max> is shown at the top of the display.

A user can print the result out by pressing

button.

The start of the next process of peak hold measurement is possible only after

removing the load from the weighing pan and pressing

button. This causes

returning to the main mode <PEAK HOLD> and automatically deleting the

pictogram <Max> on the top of display.

18.11.2. Returning to the weighing mode

CAUTION: current unit may be selected only if the weighing result is not latched.

In order to select the unit, Esc button is used. If any result has already been

latched than Esc button performs deletion of the last latched Max result.