Measured variables – Rockwell Automation 1734sc-IE4CH E+H Instruments via HART to PlantPAx User Manual User Manual

Page 134

Advertising
background image

134

Rockwell Automation Publication PROCES-UM002A-EN-P - July 2014

Appendix J

Prosonic S Transmitter

The sensor transmits ultrasonic pulses in the direction of the product surface.
There, they are reflected back and received by the sensor. The transmitter
Prosonic S measures the time t between pulse transmission and reception. From t
(and the velocity of sound c) it calculates the distance D from the sensor
membrane to the product surface:

D = c · t/2

Measured Variables

From D results the desired measuring value:

Level L
Volume V
Flow Q across measuring weirs or open channels

FDU9x

Prosonic S

FMU90

Prosonic S

FMU90

FDU9x

D

Q

100%

0%

D

L

F

E

BD

V

Item

Description

BD

Blocking distance

D

Distance from sensor membrane to fluid surface

E

Empty distance

F

Span (full distance)

L

Level

V

Volume (or mass)

Q

Flow