Measured variables, Signals from instrument to control system – Rockwell Automation 1734sc-IE4CH E+H Instruments via HART to PlantPAx User Manual User Manual

Page 90

Advertising
background image

90

Rockwell Automation Publication PROCES-UM002A-EN-P - July 2014

Appendix B

Proline t-mass 65 Thermal Flowmeter

The thermal principle operates by monitoring the cooling effect of a gas stream as
it passes over a heated transducer (PT100). Gas flowing through the sensing
section passes over two PT 100 RTD transducers, one of which is used
conventionally as a temperature sensing device, whilst the other is used as a
heater. The temperature sensor monitors the actual process values whilst the
heater is maintained at a constant differential temperature above this by varying
the power consumed by the sensor. The greater the mass flow, the greater the
cooling effect and power required in order to maintain the differential
temperature. The measured heater power is therefore a measure of the gas mass
flow rate.

Measured Variables

Mass flow
Gas temperature

Signals from Instrument to Control System

Signal

Details

Current Output

Active/passive selectable, galvanically isolated
Active: 0/4...20 mA, RL < 700

Ω (at HART: RL ≥ 250 Ω)

Passive: 4...20 mA, Ri

≥ 150 Ω, Umax = 30V DC

Pulse/Frequency Output

Active: 24 V DC, 25 mA (max. 250 mA during 20 msec), RL > 100

Ω

Passive, open collector, 30V DC, 250 mA, galvanically isolated
Frequency output: full scale frequency 2...1000 Hz (fmax = 1250 Hz), on/off ratio 1:1, pulse width max. 2s, time

constant selectable (0.0...100.0s)

Pulse output: pulse value and pulse polarity selectable, pulse width adjustable (0.5...2000 ms)

IMPORTANT

If the current output is used as a temperature output, please observe the
information for Class B according to EN 6075.

Advertising