Measured variables, Signals from instrument to control system – Rockwell Automation 1734sc-IE4CH E+H Instruments via HART to PlantPAx User Manual User Manual

Page 149

Advertising
background image

Rockwell Automation Publication PROCES-UM002A-EN-P - July 2014

149

Liquiline M CM42 Transmitter

Appendix M

Measured Variables

pH / ORP (analog sensors)

pH value
Oxidation-reduction potential (ORP)
Temperature

Conductivity (analog sensors)

Conductivity
Resistivity (conductive measurement only)
Concentration

pH / ORP (digital sensors)

Oxygen (digital sensors)

pH value
Oxidation-reduction potential
Oxygen
Temperature

Signals from Instrument to Control System

Signal

Details

Output signal

Max. load with a supply voltage of 24 V: 500

Ω

Max. load with a supply voltage of 30 V: 750

Ω

Current output 1:

1x 4...20 mA, potentially isolated against sensor circuit

Current output 1 and current output 2 (optional):

2x 4...20 mA, potentially isolated against sensor circuit

Only for HART communication:

0.8...1.2 mA peak to peak

Ex specification current output 4/20 mA

Intrinsically safe supply and signal circuit with type of protection: EEx ia IIC T6 passive

Max. input voltage Ui
Max. input current Ii
Max. input Pi
Max. internal inductivity Li
Max. internal capacity Ci

30 V
100 mA
750 mW
221

μ

H (output 1)

24

μ

H (output 2)

1.2 nF (output 1)
negligible (output 2)