Metrohm 815 Robotic Soliprep for LC User Manual

Page 31

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

2 Installation

815 Robotic Soliprep for LC

■■■■■■■■

25

5 Fasten the 6.1820.020 screw connector with the M6 connector to the

upper, larger tubing connector of the washing station.

6.1805.120

6.1820.020

6.1801.120

5

6

7

6 Connect the green 6.1805.120 FEP tubing (1 m length) to the screw

connector. This is the feed line of the washing station. Connect the
other end of the tubing to the distributor of tower 2 (see "Rinsing tub-
ings at tower 2", page 28)
.

7 Fasten a 6.1812.000 PTFE tubing to the lower tubing connector of the

washing station. This is the outlet of the washing station.

Shorten the tubing to a suitable length, so that it can be connected
to a peristaltic pump (772 Pump Unit).

Remove the union nut of lower tubing connector and guide it over
the end of a 6.1812.000 PTFE tubing. You may have to extend
the tubing end in order to be able to better mount the tubing, see
the following note.

Pull the end of the tubing over the connection nipple of the distri-
butor and fasten in place with the union nut.

Note

The opening of the tubing may need to be widened with a sharp
object (e.g. with a Phillips screwdriver).

A piece of sandpaper may be used to get a better grip on the tubing.

Do not extend the tubing end before having slid the union nut onto
the tubing.