5.4 Lift positions for stripping needles and filters

■■■■■■■■■■■■■■■■■■■■■■

48

■■■■■■■■

815 Robotic Soliprep for LC

5.4

Lift positions for stripping needles and filters

External position 1

1 Under Device selection

All devices, click on 815 Robotic

USB... and then select the tab Tower 1.

2 Fasten a plastic needle or a membrane filter to the Luer adapter of the

transfer robotic arm.

3 Under Robotic arm position actuate the button [Arrow right] until

the Luer adapter is approximately 2 cm in front of the deflector.

4 Move the lift downward until the thick sleeve of the Luer adapter is

approximately 1 cm underneath the deflector. Click on the arrow but-
ton [Arrow down] and also (if required) [Arrow up] under lift
position in order to set a suitable position.

5 Click on [Configure rack].

6 Under Lift position select Swing position External positions.

7 Click on the associated [Assign] button.

8 Under Robotic arm position select External position 1.

9 Click on the associated [Assign] button.

10 Under Lift position select the Work position for External 1.