Metrohm 815 Robotic Soliprep for LC User Manual

Page 40

Advertising
background image

2.16 Installing the Dosino

■■■■■■■■■■■■■■■■■■■■■■

34

■■■■■■■■

815 Robotic Soliprep for LC

Preparing the Dosino 1 for sample transfer

1

2

3

4

1 Screw an adsorber tube filled with cotton to the Vent connector on

the 10 mL dosing unit.

2 Equip the dosing unit with a filling tubing (Port 2 on the underside of

the dosing unit) and screw it onto a 6.1808.070 clear glass bottle (2 L,
with GL 45 thread). The bottle should be filled with the solvent (water
or an organic solvent) which is also used for sample dispersion.

3 Connect the transfer tubing to tower 1 with the end that is still free to

Port 1 on the dosing unit.

4 Place the Dosino (dosing drive) on the dosing unit.

5 Equip the connection cable of the Dosino with an anti-interference

adapter made of ferrite (T2400.102) (see drawing).

Connect the connection cable to the MSB connector 1 (MSB 1) on the
rear of the Sample Processor.