2 Installation

815 Robotic Soliprep for LC

■■■■■■■■

35

T.2400.102

6

Preparing the Dosino 2 for diluting

Install the Dosino 2 in the same way as Dosino 1. Use the 50 mL dosing unit.
Fill the clear glass bottle with a suitable dilution medium (e.g. water).

1 Screw an adsorber tube filled with cotton to the Vent connector on

the 50 mL dosing unit.

2 Equip the dosing unit with a filling tubing (Port 2 on the underside of

the dosing unit) and screw it onto a 6.1808.070 clear glass bottle (with
GL 45 thread). The bottle should be filled with the dilution medium
(e.g. water).

3 Connect the inlet tubing on the Polytron robotic arm (tower 2) with

the end that is still free to port 3 on the dosing unit.

4 Place the Dosino (dosing drive) on the dosing unit.

5 Equip the connection cable of the Dosino with an anti-interference

adapter made of ferrite (T2400.102) and connect it to the MSB con-
nector 2 (MSB 2) on the rear of the Sample Processor.