5 Setting lift positions

815 Robotic Soliprep for LC

■■■■■■■■

45

5 Under Lift position, select Work position for tower, if this is not

yet selected.

6 Click on the associated [Assign] button.

7 Click on [Close].

8 Remove the needle from the Luer adapter.

5.2

Lift position for picking up needles and filters

Picking up the needle

1 Move to a rack position 25. Under Rack position, enter Target posi-

tion 25 and click [Start].

2 Place a plastic needle in the position established in the insert.

3 Move the lift downwards until the Luer adapter engages the needle

and it is fixed in place. Under Lift position click on the arrow button
[Arrow down] in order to set a suitable position.

The needle must rest up close against the adapter. The lift may not
however be lowered too far, because otherwise the lift drive could
become overloaded and suffer damage.

4 Note the current lift position (in mm).

This lift height must be entered in the method as absolute lift position.