5.3 Lift positions for special beakers

■■■■■■■■■■■■■■■■■■■■■■

46

■■■■■■■■

815 Robotic Soliprep for LC

5 Move the lift upward and remove the needle.

Picking up the filter

1 Move to rack position 49. Under Rack position, enter Target posi-

tion 49 and click [Start].

2 Place a membrane filter on the established position in the insert.

3 Move the lift downward until the Luer adapter engages the filter and

it is fixed in place. Under Lift position click on the arrow button
[Arrow down] in order to set a suitable position.

The filter must rest up close against the adapter. The lift may not how-
ever be lowered too far, because otherwise the lift drive could become
overloaded and suffer damage.

4 Note the current lift position (in mm).

This lift height must be entered in the method as absolute lift position.

5 Move the lift upward and remove the filter.

5.3

Lift positions for special beakers

Special beaker 1

1 Move to special beaker position 1. Under Rack position, enter Target

position 103 and click [Start].

2 Place a sample beaker on the established position. Lay the cover with

the hole on the sample beaker.

3 Fasten a membrane filter on the Luer adapter of the transfer robotic

arm.

4 Move the lift downward until the filter rests against the cover of the

sample beaker. Under Lift position click on the arrow button [Arrow
down]
in order to set a suitable position.

5 Click on [Configure rack].