2 Installation

815 Robotic Soliprep for LC

■■■■■■■■

31

1

2

2

2

1

1

3

6.2740.020

6.1805.060

6.1805.120

1 Replace the three stoppers on the Polytron robotic arm with

6.2740.020 rinsing nozzles.

2 Connect the three rinsing tubings that are connected to the tower 2

distributor to the rinsing nozzles.

3 The preinstalled tubing of the robotic arm is used to add solvent to the

sample. Connect a 6.1805.120 to it and connect the other end of the
tubing to to the port 1 of the Dosino with a 50 mL dosing unit.

2.14

Connecting the Polytron

®

The Polytron

®

PT 1300 D is comprised of a control module and a disper-

sion drive with a dispersion aggregate for mixing and shredding solid
samples. The control device is directed by a PC software, e.g. tiamo via a
serial RS232 connector.

Consult the Polytron user manual for details about setting up the connec-
tions. The Polytron is connected as follows:

1 Connect the connection cable of the dispersion drive to the front side

of the control module.

2 Check the mains voltage setting on the connection socket on the rear

side of the control module. Switch the device off. Plug in the mains
cable and connect with a socket.

3 Connect the 6.2134.040 RS232 connection cable to the 9-pin con-

nection socket on the rear side of the control module. Plug in the other

Advertising