Metrohm 815 Robotic Soliprep for LC User Manual

Page 35

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

2 Installation

815 Robotic Soliprep for LC

■■■■■■■■

29

1 Open the guide chain

Insert a screwdriver into the groove located on the side of a chain
link.

Loosen the clip with a forceful leverage movement.

Pull the clip out of the chain by hand.

Repeat the above actions for each chain link.

2 Insert into the guide chain

Place the required tubings and cables into the guide chain.

3 Close the guide chain

Close the clip for each chain link again by hand and apply forceful
pressure to snap them into place.

Caution

Take care to ensure that any cables and tubings that are fixed in the guide
chain are not able to hinder the free movement of the lift or of the robotic
arm. This could lead to damage to the affected drive.

The clip of one chain link can be removed entirely if required by releasing it
on both sides.

Advertising