5 Setting lift positions

815 Robotic Soliprep for LC

■■■■■■■■

51

Work position

1 Move the lift upwards by several mm.

2 Fasten a membrane filter to the Luer adapter of the robotic arm.

3 Under Lift position actuate the arrow key [Arrow down] several

times until the membrane filter is tightly placed on the Luer connector.

4 Click on [Configure rack].

5 Under Lift position select the Work position for External 3.

6 Click on the associated [Assign] button.

7 Click on [Close].

8 Move the lift to the Shift position and remove the membrane filter

from the Luer adapter.

5.6

Lift position for the Polytron

Working height

1 Under Device selection

All devices, click on 815 Robotic

USB... and then select the tab Tower 2.

2 Move to sample position 1. Under Rack position, enter Target posi-

tion 1 and click [Start].

3 Move the lift downward until the Polytron is at a suitable height for

sample dispersion. Click on the arrow button [ Arrow down] and also
(if required) [Arrow up] under lift position in order to set a suitable
position.

4 Click on [Configure rack].

5 Under Lift position, select Work position for tower, if this is not

yet selected.