1.3 Safety instructions

■■■■■■■■■■■■■■■■■■■■■■

6

■■■■■■■■

864 Robotic Balance Sample Processor

Warning

Always install the safety shield supplied with the equipment before
using the instrument for the first time. Pre-installed safety shields are
not allowed to be removed.

The 864 Robotic Balance Sample Processor may not be operated with-
out a safety shield!

Warning

Personnel are not permitted to reach into the working area of the
instrument while operations are running!

A considerable risk of injury exists for the user.

Warning

In the event of a possible blockage of a drive, the mains plug must be
pulled out of the socket immediately. Do not attempt to free jammed
sample vessels or other parts while the device is switched on. Blockages
can only be cleared when the instrument is in a voltage-free status; this
action generally involves a considerable risk of injury.

Warning

The 864 Robotic Balance Sample Processor is not suitable for utilization
in biochemical, biological or medical environments in its basic equip-
ment version.

Appropriate protective measures must be implemented in the event
that potentially infectious samples or reagents are being processed.