Metrohm 864 Robotic Balance Sample Processor User Manual

Page 45

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

864 Robotic Balance Sample Processor

■■■■■■■■

37

3 Prepare the titrant

Fasten a filling tubing and an adsorber tube (filled with wadding)
on a 6.3032.220 20 mL dosing unit.

Screw a dosing unit on a bottle with titrant.
A 6.1608.030 clear glass bottle (1 L, with GL45 thread) can be
used.

On Port 1 of the dosing unit, fasten one of the 6.1805.130 FEP
tubings
(100 cm), which are already mounted on the titration
head, see previous chapter.

4 Mount the drive for the titrant

Attach an 800 Dosino drive to the dosing unit.

Connect the connection cable to the connector MSB 1 of the
Titrando.

5 Prepare the solvent

Fasten a filling tubing and an adsorber tube (filled with wadding)
on a 6.3032.250 50 mL dosing unit.

Screw a dosing unit onto the 6.1608.070 clear glass bottle (2 L,
with GL45 thread) filled with solvent.

On Port 1 of the dosing unit, fasten one of the 6.1805.130 FEP
tubings
(100 cm), which are already mounted on the titration
head, see previous chapter.

6 Mount the drive for the solvent

Attach an 800 Dosino drive to the dosing unit.

Connect the connection cable to the connector MSB 2 of the
864 Robotic Balance Sample Processor.

7 Connecting the electrode cable

Connect the 6.2104.030 electrode cable which is already con-
nected to the electrode to the connector Ind. on the rear of the
Titrando.

8 Connect the Titrando

Connect a 6.2151.000 connection cable to the Titrando (Con-
troller
connector).

Connect the cable to a USB connector on the Sample Processor
or on the PC.

After that, the Titrando is recognized by the tiamo

™ software and

can be added to the device table. The same applies for the dosing

Advertising