3.13 Installing the balance and the ionizer

■■■■■■■■■■■■■■■■■■■■■■

40

■■■■■■■■

864 Robotic Balance Sample Processor

3 Connect the cables

Connect the ground cable for the balance to the metal clip at the
underside of the housing of the Sample Processor.
The clip is located on the rear part of the bottom of the housing.

Connect the RS-232 connection cable of the balance to the
COM1 connector of the PC.

Connect the mains cable with the mains adapter of the mains
supply.

Note

Afterwards, check once again to see if the balance is still leveled.
For this, use the supplied 6.2831.000 inspection mirror. This
makes it possible to read off the spirit level of the balance, even if
it is now located under the Sample Processor.

Setting the interface parameters

The parameters of the data transmission between balance and PC must
match on both devices.

We recommend using the following parameters:

Baud rate 9600

Data bit 8

Parity None

Stop bit 1

Handshake None

1 Set the RS-232 parameters of the balance

Setting the parameters above is described in the manual of the bal-
ance.

2 Register the balance in tiamo

In the tiamo

™ software, click the symbol Configuration.

In the device table click [Edit] and select New

Under Balances, select the balance type Precisa and click [OK].
The properties window of the balance is displayed.

Enter the serial number of the balance.
This is necessary to unambiguously identify the device.

3 Set the RS-232 parameters

Click the tab RS 232.