3 Installation

864 Robotic Balance Sample Processor

■■■■■■■■

41

Under COM Port, select the connector on the PC the connection
cable was connected to.

Enter the values listed above as RS-232 parameters.

Click [Connect].
The dialog window Establish connection is displayed.

Press the [Print] key on the balance.
The data sent by the balance is displayed.

Close the dialog window with [OK].
The input fields in the properties window are now locked. This
means that the connection between balance and PC has been
established.

Close the properties window with [OK].

Installing the ionizing bar

In order to eliminate electrostatic loads while weighing, an ionizing bar
must be installed as close to the balance as possible. Proceed as follows:

6.2057.160

1

2

3

Figure 26

Installing the ionizing bar

1 Fasten the holder

Fasten the 6.2057.160 holder on the assembly rail with the aid of
a hexagon key.
The position must be selected in such a way that the recess of the
holder surrounds the weighing column.

2 Insert the screws

Insert the two accompanying screws from above into the holding
rail of the ionizing bar.