3 Installation
864 Robotic Balance Sample Processor
■■■■■■■■
55
3.19
Adjusting the lift positions
Fine adjustment
Some lift positions must be adjusted precisely in order to ensure a perfect
automation sequence. Use the manual control of the instrument for this
purpose. The following section describes how you can approach the indi-
vidual positions and adjust them as needed.
1 Open the manual control
■
In the sidebar of tiamo
™, click the hand symbol.
■
In the left-hand window, under 864_1 (864 Balance …, click
the item Tower 1 and then select the Move tab.
Lift position for picking up pipetting tips
1 Move to position
■
Under Rack position, enter Target position 21 and click
[Start].
■
Place a pipetting tip on the position moved to.