1 connecting dosing devices, Connecting dosing devices – Metrohm 864 Robotic Balance Sample Processor User Manual

Page 68

Advertising
background image

3.21 Connecting MSB devices

■■■■■■■■■■■■■■■■■■■■■■

60

■■■■■■■■

864 Robotic Balance Sample Processor

Caution

Exit the control software before you plug MSB instruments in. The con-
trol instrument recognizes when it is switched on which instrument is
connected at which MSB connector. The operating unit or the control
software enters the connected MSB devices into the system configura-
tion (Device manager).

MSB connections can be extended with the 6.2151.010 cable. The length
of the connection must not exceed a maximum of 15 m.

3.21.1

Connecting dosing devices

Three dosing devices can be connected to the instrument.

The types of dosing devices that are supported are:

800 Dosino

700 Dosino

805 Dosimat

685 Dosimat

Warning

If a Dosino is connected to the 864 Robotic Balance Sample Processor,
then the connection cable must be equipped with a ferrite core
T.2400.102. The ferrite core reduces any interference voltages that may
occur and thus ensures compliance with strict EMC standards pursuant
to applicable technical norms, see Chapter "Technical Data".

Proceed as follows:

1 Mounting ferrite core

Fasten a ferrite core T.2400.102 to the Dosino connection cable near
to the plug.

2 Connect a dosing device

Exit the control software.

Connect the connection cable to one of the sockets marked with
MSB on the rear of the control instrument.

Start the control software.

Advertising