Metrohm 864 Robotic Balance Sample Processor User Manual

Page 29

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

864 Robotic Balance Sample Processor

■■■■■■■■

21

Tower 1

1 Open the robotic arm configuration

Click the Configuration icon.

Double-click the instrument name 864_1 in the Devices win-
dow.

In the properties window of the 864 Robotic Balance Sample Pro-
cessor under Tower 1, open the configuration of the Swing
Head
.

2 Enter the configuration data

Enter the following data:

– Swing offset –8.6°
– Maximum swing range 122°
– Swing radius 149.8 mm
– Rotation offset 0.0°
– Swing direction

Close the configuration window with [OK].

Tower 2

1 Open the robotic arm configuration

In the properties window of the 864 Robotic Balance Sample Pro-
cessor under Tower 2, open the configuration of the Swing
Head
.