Metrohm 864 Robotic Balance Sample Processor User Manual

Page 56

Advertising
background image

3.17 Adjusting the rack and the robotic arm

■■■■■■■■■■■■■■■■■■■■■■

48

■■■■■■■■

864 Robotic Balance Sample Processor

The last position on the sample rack is conceived as Adjusting posi-
tion
.

3 Define the special beaker on the 6.2068.010 sample rack

Select the lowest line (Special beaker 16) and click [Edit].

Under Rack position, select 61.

Close the dialog window with [OK].

Close the rack data table with [OK].

Moving to the positioning reticle

Proceed as follows:

1 Open the manual control

In the Manual control in tiamo

™, click the tab Move.

2 Move to special beaker 16

Under Rack position, select the target position Special beaker
16
and click [Start].

It is also possible to specify the target position as absolute rack posi-
tion. For the 6.2068.010 sample rack, the adjusting position is the
no. 61.

3 Move to the adjusting position

Unscrew the FEP tubing from the adapter on the robotic arm.

Insert the PEEK capillary supplied from above into the adapter.
The capillary must protrude below from the adapter by around
0.5 cm (see following illustration)

Under lift position enter 100 mm as target position and click
[Start].