5 troubleshooting, 1 robotic arm, 2 balance – Metrohm 864 Robotic Balance Sample Processor User Manual

Page 75: Robotic arm, Balance

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Troubleshooting

864 Robotic Balance Sample Processor

■■■■■■■■

67

5 Troubleshooting

5.1

Robotic arm

Problem

Cause

Remedy

The robotic arm
moves all the way
outward and buz-
zes.

Sample Processor – The
Swing Head is not correctly
configured.

In the control software under "Configuration"
(or under "Device manager" for Touch Con-
trol), enter the correct value for the Swing
offset
.

Sample Processor –
Robotic arm is wrongly
mounted.

Disconnect the mains plug and dismount the
robotic arm. Check the configuration of the
robotic arm and mount it correctly if necessary
(left-swinging

⇔ right-swinging).

The Swing Head
either misses the
rack positions
totally or is inaccu-
rate

Sample Processor – The
Swing Head is not correctly
configured.

In the control software under "Configuration"
(or under "Device manager" for Touch Con-
trol), enter the correct values for the Swing
radius
, Swing offset etc.

Sample Processor – The
axial distance is not cor-
rectly configured.

In the control software under "Configuration"
(or under "Device manager" for Touch Con-
trol), enter the correct value for the Axial dis-
tance
.

Sample Processor – The
wrong rack table is being
used.

Initialize the rack using the function Initialize
rack
in the "Manual control".

Swing Head – The Swing
Head drive is defective.

Contact the Metrohm Service.

5.2

Balance

Problem

Cause

Remedy

The sample size is
not or not correctly
received.

The parameters of the
RS-232 interface are
wrongly set.

Correct the RS-232 parameters (see page
39ff)
. They have to match with those of the
balance.

The connection cable is
unsuitable.

Use the RS-232 cable of the balance manufac-
turer or ask for the corresponding cable from
Metrohm.

Advertising