Figure 34, Centering the weighing column – Metrohm 864 Robotic Balance Sample Processor User Manual

Page 60

Advertising
background image

3.18 Adjusting the weighing equipment

■■■■■■■■■■■■■■■■■■■■■■

52

■■■■■■■■

864 Robotic Balance Sample Processor

Unscrew the weighing plunger and dismantle the drip pan.

Screw the weighing plunger back on again and attach the rack.

Move to a free sample position on the outermost row of the rack
(target position Tower 1) with the aid of the control software
(tiamo

™).

Remove the weighing pan from the sample position that is moved
to.

Centering the weighing column

1

2

2

Figure 34

Centering the weighing column

Proceed as follows:

1 Loosen the clamping ring

Loosen the clamping ring of the weighing column.

2 Position the balance

Shift the position of the balance in such a way that the weighing
plunger is located precisely in the middle of the opening of the
sample rack.

3 Check the leveling

Check the horizontal position of the balance with the 6.2831.000
inspection mirror and correct it as necessary. The spirit level for
the optical inspection is located on the rear side of the scale, i.e.
underneath the housing of the Sample Processor.

4 Fix

Fix the clamping ring again.