Figure 15, Mount the robotic arm with the titration head – Metrohm 864 Robotic Balance Sample Processor User Manual

Page 36

Advertising
background image

3.6 Setting up the Swing Heads and robotic arms

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

864 Robotic Balance Sample Processor

1

2

3

6.1458.040

6.1462.260

Figure 15

Mount the robotic arm with the titration head

1 Mount the titration head insert

Place the 6.1458.040 titration head insert in the opening of the
robotic arm and screw tight with the supplied screws.
The arrow marking on the edge of the insert must be pointed
towards the Swing Head.

2 Position the robotic arm

Hold the robotic arm in such a way that the titration head faces
left and slip it over the guide pins of the drive disc from below.

While doing so, rotate the robotic arm outwards as far as possi-
ble, i.e. towards the tower - see previous figure.

Note

Take care to ensure that you do not twist the drive disc, thus caus-
ing pressure against the drive.

3 Fix the robotic arm

Screw the robotic arm to the Swing Head tightly with the screws
and washers provided.