2 principle of data input – Metrohm 787 KF Titrino User Manual
Page 11
2.2 Principle of data input
787 KF Titrino
7
2.2 Principle of data input
• The first line of the display always shows the active
mode (in this case "KFT") and the state of the Titrino.
"********" means the Titrinos is in its standby state.
• The keys <CONFIG>, <PARAM> and <CALC
DATA> are opening the corresponding menus.
• Pressing one of this keys shows the first group of
inquiries of the corresponding menu in the second
line of the display.
Example key <CONFIG>:
You opened the configuration menu, the second line
of the display shows the first group of inquiries of
this menu ">KF device settings".
• By pressing <CONFIG> repeatedly you can move
to the other titles of this menu. With pressing
<CONFIG> 2 times you go to the group of inquiries
"peripheral units".
• If a dialog text is marked with ">", it contains a
group of inquiries. You branch into this group by
pressing <ENTER>.
• Example first inquiry of the group "peripheral units":
Now you opened the first inquiry of the group pe-
ripheral units in the configuration menu and you can
change the setting.
If a dialog text of an inquiry is marked with ":", you
can select a value with the key <SELECT>.
• A value is stored with <ENTER> and the cursor
moves to the next inquiry.
• Repeated pressing of <ENTER> moves you
through the inquiries of the group ">peripheral
units“, after the last inquiry of this group you leave
this group and return to the level above.
The next group of the menu "configuration“ appears:
">auxiliaries“
• With key <QUIT> you leave an inquiry or a group of
inquiries, it always moves you one level up.
• In this example you leave the menu "configuration"
by pressing <QUIT> and return to the standby
state with the display of the titration mode.
KFT ********
send to:
IBM
KFT ********
>auxiliaries
KFT ********
KFT ********
>KF device settings
KFT ********
KFT ********
>peripheral units