Metrohm 787 KF Titrino User Manual

Page 17

Advertising
background image

2.3

Tutorial

787 KF Titrino

13

Viewing the statistics data

If you have not connected a printer, you can view the statistics data on the display with

<SELECT>.

<SELECT>

<SELECT>

<SELECT>

<SELECT>

• From the result display "titer", switch to the display of

the mean value with <SELECT>.

Mean value of 2 determinations.
• Press <SELECT> again and the standard deviation

is shown.

Absolute standard deviation.
• Pressing <SELECT> once more leads to the

relative standard deviation.


• <SELECT> returns you to the result display.

Repeat the titer determination once more.

With a connected printer you may want a printout of the single results that are used for the

statistics calculation: press the key sequence <PRINT><1><ENTER>.

Let us assume that you have discovered that one of the values is an "outlier" and you

would like to delete it.

Deleting a result in the mean value calculation

2 × <CALC DATA>

2× <ENTER>

<SELECT>

<ENTER>

In our example, the deviation of the second result is

excessive. We thus wish to delete it:
• Press <CALC DATA> twice. The display shows:

• Open with <ENTER> the inquiry of the result table

"res.tab:"

• Press <SELECT> to display "delete n". This means

that you wish to delete a single result with the index

n in the result table.

KFR volume

5.676 ml

titer 5.3686

mg/ml

KFR volume

5.676 ml

res.tab: original

KFR volume

5.676 ml

res.tab: delete

n

KFR volume

5.676 ml

mean (2)

5.3477 mg/ml

KFR volume

5.676 ml

+/-s 0.02973

mg/ml

KFR volume

5.676 ml

s(rel) 0.56

%

KFR volume

5.676 ml

>statistics

Advertising