Metrohm 787 KF Titrino User Manual

Page 39

Advertising
background image

2.8 Result calculations

787 KF Titrino

35

Statistics calculations

If the mean value is active, the following quantities are calculated:

• mean
• standard deviation
• relative standard deviation

These values are printed out in the result report or they can be viewed with the
<SELECT> key:

Press the <SELECT> key repeatedly until the desired value appears in the display. The

following are displayed in succession, see also page 13.

• current result
• mean
• standard deviation
• relative standard deviation


A table of the single results can be printed out with the key sequence
<PRINT><1><ENTER>.

Remarks:

• If the number of single results n has been reached, the statistics calculation

begins anew at the next start.

• When the mode is changed, the statistics instructions of the previous mode come

to an end (res.tab. deleted) and the instructions of the new mode are followed.

• Old results that are no longer needed for the statistics calculation can be deleted

with "delete all".

• Recalculated single results are immediately incorporated in the statistics

calculation.

• If a result can not be calculated, the run counter is incremented. The counter for

the number of reconciled single results remains unchanged, however.

Advertising