Metrohm 787 KF Titrino User Manual
Page 28
2.6. Parameters, key <PARAM>
787 KF Titrino
24
>preselections
conditioning: OFF
req.ident: OFF
req.smpl size: OFF
report: OFF
cond.
Preselections for the sequence
Automatic conditioning of titration vessel. (ON, OFF)
If conditioning is "on", between the titrations the titration
solution is constantly maintained at the end point.
When conditioning is performed, the volume drift can
be displayed during the conditioning with <DRIFT>,
see page 36.
Request of the sample identification after start of
titration (ON, OFF)
After start, the sample identification can be requested
automatically.
Not performed or inquired in the "BLANK" mode.
Request of sample size after start of titration (ON, OFF)
Not performed or inquired in the "BLANK" mode.
Selection of the result report on titration completion
(short, full, OFF)
See also page 38.