5 appendix, 1 technical specifications – Metrohm 787 KF Titrino User Manual

Page 67

Advertising
background image

5.1 Technical specifications

787 KF Titrino

63

5 Appendix

5.1 Technical specifications

Modes

KFT:

Karl Fischer titration

TITER:

Titer determination with water or water-

containing standard, resp. with sodium tar-

trate

BLANK:

Blank determination

Measuring input

1 measuring input for polarized electrodes.

Measuring range

Voltage

0...±1500 mV, resolution 1 mV, error limit 0.1 % fullscale

Current

0...±150

µA, resolution 1 µA

Polarizer
Ipol

0...±127

µA, resolution 1 µA

Upol

0...±1270 mV, in steps of 10 mV

Measuring amplifier (at 25 °C and Titrino warmed-up)
Input

resistance

>10

13

Offset

current

<3 ∗10

–13

A

Deviation of offset voltage

15 µV/K

Dosification

Volume of buret cylinder 1, 5, 10, 20 or 50 mL

Resolution

10 000 steps per buret cylinder

Titrating buret

1 internal buret

Auxiliary buret

additional: 776 or 765 Dosimat

Materials

Housing

Polybutyleneterephthalate

(PBTP)

Keypad cover

Polycarbonate (PC)

Display

LCD, 2 lines of 24 characters each

Height of characters 5 mm

LED

back-lit

Advertising