2.3. Tutorial

787 KF Titrino

12

Determination

<START>

0.03

<ENTER>

• Now prepare the sample, e.g. 30 µL water and start

the titration with <START>. The display shows

• Inject the sample and enter the weight in g:

If you have connected a balance, you can also back-

weigh the syringe.

The titration now runs.

The first line of the display shows the current volume. In
addition, a "↑" appears when dispensing is being per-

formed.

A "control bar" appears on the second line and shows

the control deviation of the current measured value

from the set endpoint.

On completion of the titration, the first line shows the

dispensed volume and the second the calculated titer.

If you have connected a printer, a report is also printed:

Perform another determination.

On completion of the titration, the mean value of the two determinations is calculated. The

report on the printer has the following appearance:

KFR volume

0.000 ml

smpl size

1.0 g

KFR volume 1.426

ml

#======

KFR volume

5.632 ml

titer 5.3267

mg/ml

date 2002-03-27 time 13:55:10 3

smpl size 0.03 g

titer 5.3267 mg/ml

============

date 2002-03-27 time 13:55:10 3

smpl size 0.03 g

titer 5.3686 mg/ml

mean (2) 5.3477 mg/ml

+/-s 0.02973 mg/ml

s(rel) 0.56 %

============