Advertising
background image

3.1 Error and special messages

787 KF Titrino

39

3 Error messages, Troubleshooting

Data transfer inoperative See measures on page 41.

3.1 Error and special messages

change solvent

The limit volume of the reagent volume counter "limit KF reag."

has been reached..

Exit: Change solvent and press <CLEAR>.

check electrode

There is a break or short circuit. Possible causes and

rectification of the fault:

- the electrode is not plugged in ⇒ plug it in

- the electrode is not immersed in the solution ⇒ immerse it

- the electrode is defective ⇒ use new electrode.

- the electrode cable is defective ⇒ use new cable.

The electrode test can be switched off under the <PARAM>

key.

Exit: Rectify fault or <STOP>.

check exchange unit

The Exchange Unit is not mounted (properly).

Exit: Mount Exchange Unit (properly) so that the coupling

engages or <STOP>.

cylinder empty!

During dispensing with <DOS> an entire cylinder ahs been

dispensed.

Exit: Press <STOP/FILL>.

division by zero

The result could not be calculated as a divisor in the formula

was equal to zero.

Exit: Enter appropriate value.

stop V reached

The determination has been stopped as the stop volume has

been reached.

system error 3

The instrument adjustment data have been overwritten.

Exit: <CLEAR>. Default adjustment data are set. The error

message appears each time the instrument is switched on

until it has been readjusted (Metrohm service).

????

The result exceeds the valid data range or too many decimal

places are needed to display.