2.5 Configuration, key <CONFIG>
787 KF Titrino
21
>auxiliaries
dialog: english
date 2001-03-27
time 08:13
run number 0
electrode test: ON
display KFR vol.: ON
device label
program 787.0010
General settings
Selection of dialog language (english, deutsch,
francais, español)
Current date (YYYY-MM-DD)
Format: Year-month-day, entry with leading zeros.
Current time (HH-MM)
Format: Hours-minutes, entry with leading zeros.
Current run number for result output (0...999)
The sample number is set to 0 when the instrument is
switched on and incremented on every determination.
Electrode test (ON, OFF)
Test for polarized electrodes. Performed on
changeover from the inactive state to a measurement.
"OFF" means that the test is not performed.
Display the dispensed KFR volume in the result display
(ON, OFF)
"ON" means the dispensed KFR volume will be
displayed in the result output on the first line.
Individual identification of devices (up to 8 ASCII
characters)
Display of program version