2.6. Parameters, key <PARAM>

787 KF Titrino

22

2.6 Parameters, key <PARAM>

The key <PARAM> is used for the entry of values that

determine the titration sequence. Values marked with

"cond." are accessible during the conditioning. "**titr."

means that these values can be changed during the

titration. They influence the ongoing determination.

Other values can only be changed in the inactive state.

The display texts of the Titrino are shown to the left.

The values are the default values.

2.6.1 Parameters








titration parameters

are valid for the global titration sequence.

preselections:

ON/OFF of various auxiliary functions such as auto-

matic requests after the start and report.


>titration parameters

extr.time 0 s
**titr.







stop crit: drift
**titr.
stop drift 20 µl/min
**titr.

t(delay) 10 s
**titr.



stop V 99.99 ml
**titr.


start V: 0.00 ml
cond.

General titration parameters

Extraction time (-9999.. 9999 s)

for

-

that evolve sample slowly

-

work with KF oven

With positive extraction times, titration is performed

continuously during this time; with negative values no

titrant is added during the extraction time. The

extraction time can be aborted with <QUIT>.

Type of stop criteria (drift, time)

Titration stops if EP is and stop drift is reached. (1...999

µL/min)

Titration stops if there is no dosing during t(delay).

(0...99 s)

Switch off when the end point is reached and the set

time after the last dispensing has elapsed.

Stop volume (0...99.99 mL, OFF)

Safety stop, e.g. if no endpoint is found.

<CLEAR> sets "OFF".

Start volume (0.00…99.99 mL.)

Performed and inquired only in KFT mode.

Start volume is dispensed before titration start. In the

PARAM

KFT ********

>titration parameters