4.1 Setting up and connecting the instruments

787 KF Titrino

57

4.1.4 Connection of a KF Oven

It is expedient to place the oven on 6.2041.180 instrument bridge.

Take care that the gas outlet of the oven enters the titration vessel as

directly as is possible to prevent the formation of condensed water in

the outlet tubing.

The 707 KF Oven can also be connected instead of the 768 KF

Oven.

4.1.5 Connection of the 774 Oven Sample Processor

The Oven Sample Processor heats the sample and transfers the

moisture from the sample to the titration vessel of the Titino. Titrino

and Oven Sample Processor are connected via the remote sockets

(cable 6.2141.020).

774


• The "Remote" socket allows not only connection of a sample

changer but also additional control functions. Pin assignment of

the "Remote" socket and control possibilities, see page 65f.

Cable (Remote) 6.2141.020

C

B

A

D

787

Cable (Remote) 6.2141.010

768

A

B

C

D

C

B

A

D

787