2.6 Parameters, key <PARAM>

787 KF Titrino

23




dos.rate max. ml/min
**titr.










max.rate 10.0 ml/min
**titr.










min.volume incr. min. µl
**titr.




case of a negative extraction time, it is dispensed after

the extraction time has elapsed.

If a start volume has been entered, an inquiry appears

regarding the dosing rate:

Dosing rate for start volume

(0.01...150 mL/min, max.)

<CLEAR> sets "max.".

The maximum rate depends on the Exchange Unit:

Exchange Unit max.

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

Maximum titration rate (0.01...150 mL/min, max.)

<CLEAR> sets "max.".

This parameter determines primarily the addition rate

outside the control range, see also page 26.

The maximum rate depends on the Exchange Unit:

Exchange Unit

max.

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

Minimum volume increment in the titration (0.01...999.9

µL/min, min.)

<CLEAR> sets "min.".

This parameter determines the addition rate right at the

start and the end of the titration, see also page 26.

This parameter influences the titration speed and

therefore its accuracy very strongly: A smaller min.rate

results in a slower titration.

The minimum increment depends on the Exchange

Unit:

Exchange Unit

min.

5 mL

0.5 µL

10 mL

1.0 µL

20 mL

2.0 µL

50 mL

5.0 µL