2.5 Configuration, key <CONFIG>

787 KF Titrino

19

2.5 Configuration, key <CONFIG>

Key <CONFIG> serves to enter device specific data.

The set values apply to all modes. The quantities

marked with "cond." are also accessible during the

conditioning

KF device settings:

Settings for the monitoring of the reagent supply, the

measuring input, the endpoint and the filling rate.

peripheral units:

Selection of printer, balance and the curve at the ana-

log output.

auxiliaries:

e.g. setting of dialog language, date, time, etc.

RS232 settings:

RS parameters for the COM interface.

The display texts of the Titrino are shown to the left.

The values are the default values.


>KF device settings

limit KF reag. OFF ml








actual KF reag. 0 ml

polarizer: I(pol)

I(pol) 50 µA
EP 250 mV
cond.


U(pol) 400 mV
EP 25 µA
cond.








KF device settings for all modes

KF reagent volume counter (0...999 mL, OFF)

is used for monitoring of the
• reagent supply

• the buffer capacity of the solvent with 2-component

reagents

<CLEAR> sets "OFF".

If the KF reagent volume counter is activated, the

inquiry of the current counter status follows.

Current counter status (0...999 mL)

Selection of the polarization mode (I(pol), U(pol))

Polarization current(-127...127 µA)

Endpoint voltage(-1500 mV...1500 mV)

Polarization voltage(-1270...1270 mV, in steps of 10 mV)

Endpoint current (-150...150 µA)

CONFIG

KFT ********