Metrohm 787 KF Titrino User Manual

Page 38

Advertising
background image

2.8. Result calculations

787 KF Titrino

34

Formula for the blank determination BLANK

factor

)

KFR

(

volume

blank

×

=

If the blank is determined with an excessive solvent volume, the result can be

converted to the amount used subsequently with the aid of the factor.


If the drift correction is switched on, the "volume(KFR)" is corrected in all modes as
follows:

volume(KFR) – d = volume(KFR) – (drift value * (-d)time)


"(-d)time" corresponds to the time during which control is exerted (i.e. during the positive
extraction time and during the titration time, but not during the dispensing of the start

volume and during negative extraction times).
With manual drift correction, the entered drift value is reconciled in the result calculation,
whereas with automatic drift correction the drift value at the start of the titration is used.

The drift correction is used when the ingress of moisture during the titration needs to be

compensated. A drift correction is appropriate when the ratio


is large.

Key <UNIT>





The <UNIT> key can be used to select the unit of the

result in the KFT mode and of the sample size. It is

accessible during the conditioning and in the standby

state.

In the modes "TITER" and "BLANK", the result unit can

be viewed only but not changed.



The display texts of the Titrino are shown to the left.
The values are the default values.

result unit: %


result unit: %;2




smpl size unit: %

Unit of the result in the KFT mode (%, ppm, mg/mL, g,

mg, mL, mg/pc, no unit)

Number of decimal places (0...9)

The number of decimal places in the result output

follows the separator ";". The number flashes to draw

attention to the fact that this value can now be entered.

Unit for the sample size (g, mg, mL, µL, pc, no unit

)

drift ∗ titration time

volume(KFR)

UNIT