3.3. Diagnosis

787 KF Titrino

48

9 RS 232 test

A 3.496.8480 test plug normally used in the repair service is required for this test. However, this plug can also be

purchased by customers under the above number.

For the sake of completeness, the procedure is described here. If a diagnostic test of the RS 232 interface is not re-

quired, continue with point 10.

Fig. 4

Connections in the 3.496.8480 plug

• Press <5>.

RS232 test

• Press <ENTER>.

RS232 test-connector?

• Insert the 3.496.8480 test plug in 'RS 232' port.

• Press <Enter>.

Test

sequence:

1. In case of an error an error message is displayed. If for example no test plug is con-

nected, error message

error 68

appears).

2. If no error is found, after about 5 s display shows

RS232 test o.k.

• Remove test plug.

• Press <Clear>.

diagnose press key 0...9

TxD

RxD

DCD

DTR

DSR

RTS

CTS

2

3

8

20

6

4

5

Advertising