Metrohm 787 KF Titrino User Manual

Page 36

Advertising
background image

2.8. Result calculations

787 KF Titrino

32

divisor 0.1

cond.




blank 0.0 ml

cond.





drift corr.: auto

cond.











drift value 0.0
µl/min

cond.

Divisor (

±

1000000)

in the denominator of the calculation formula. Used

mainly for the density, see page 33.

Request only in the "KFT" mode.

Blank (0.0...99.9991 mL)

The blank value is written in automatically from the

blank determination mode. Values that are corrected

by a blank value are flagged with "-b".

Request only in the "KFT" mode.

Drift correction for the result (auto, man., OFF)

With the drift correction switched on, the value

(drift*titration time) is subtracted from the KFR volume.

Values that has been corrected by the drift are flagged

with "-d".

auto: The drift value at the start of the titration is

subtracted.

man: A fixed value of the drift is entered and

subtracted.

If the drift correction is set to "man.", the drift value is

requested:

Drift value for the manual drift correction (0...99.9

µL/min)

>statistics

mean n= 20

cond.

res.tab: original

cond.






delete n= 1

cond.

Statistics calculation

Mean value calculation from n single results (2...20)

Result table (original, delete n, delete all)

"original":

The original table is used. Deleted

individual results are again incorporated in

the evaluation.

"delete n": Deletion of single results with the index n.

"delete all": The entire table is deleted.

Delete data from sample number n (1...20)

The deleted result is removed from the statistics

calculation.