2.10. Data output

787 KF Titrino

38

Example of reports:

Full result report

- date, time, current run number.
- sample size, identification.

- The blank is printed out only if the value is ≠ 0 mL.


- The drift and the (-d)time are printed out only it the drift

correction has been set to "OFF".

- result

- statistics



Short result report

- date, time, current run number.


- result

- mean value with 5 single results
- standard deviation
- relative standard deviation

Additional possibilities for report outputs
In addition to the reports which are printed at the end of the titration, various other reports

can be put out. There are 2 possibilities to select the reports:

<PRINT><keyX><ENTER>

key X is the key under which the appropriate data

are entered.

List of reports:

Report

<Key X>

Configuration report

CONFIG

Parameter report

PARAM

Calculation data

CALC DATA

Table of the individual results for statistics

calculation

1

Result display without a printer
Results can also be viewed in the display. After a determination the calculated result and
the single results of the statistics calculations can be displayed with <SELECT>.

<SELECT> switches to the next result, view also page 13.

date 2002-03-27 time 08:33:42 4
smpl size 1.0 g ident. 08-4
KFR volume 2.278 ml
blank 0.0315 ml
titer 5.0 mg/ml
drift auto 1.9 µl/min

(-d)time 0:43
water 1.14 %
mean(4) 1.13 %
+/-s 0.005 %
s(rel) 0.44 %
============


date 2002-03-27 time 08:35:42 5
smpl size 1.0 g ident. 08-5

(-d)time 0:45
water 1.14 %
mean(5) 1.13 %
+/-s 0.005 %
s(rel) 0.44 %
============

Advertising