2.3

Tutorial

787 KF Titrino

17

Drift correction of the results
You can correct your results by the drift. This is primarily necessary if you wish to deter-

mine very small amounts of moisture or when you have very long titration times.
For the correction "drift value ∗ titration time" is subtracted form the dispensed volume:

corrected volume = dispensed volume – (drift value ∗ titration time)

The drift value can be determined automatically or entered manually. In the automatic

determination the current drift value at the start of the titration is stored and used for the

correction. In the manual entry you yourself enter a fixed value, e.g. the stable drift value

that you have read off from the display during conditioning.

<CALC DATA>

7 × <ENTER>

2 × <SELECT>

<ENTER>

2.2

<ENTER>

<QUIT>

• Activate the drift correction: Press <CALC DATA>

and enter the inquiries of the "calculation" group with

<ENTER>. Proceed with <ENTER> until the dis-

play shows:

• Make your selection with <SELECT>, e.g. "man.".

• Enter a drift value, e.g. 2.2 µL/min and exit the in-

quiry.

The result is recalculated and the first line of the display

shows

-d means that a drift correction has been performed.

If you have connected a printer, the full report provides

you with all data for recalculation of the result:

KFT volume

3.459 ml

drift corr.:

OFF

KFT volume

3.459 ml

drift value

0.0 µl/min

KFT-Vol -d

2.365 ml

date 2002-03-27 time 13:55:10 3

smpl size 0.03 g

KFR volume 3.459 ml

drift man. 2.2 µl/min

(-d)time 1.03

water 922.2 mg/ml

============