2.8 Result calculations

787 KF Titrino

31

2.8 Result

calculations

Key <CALC DATA>

The key <CALC DATA> is used to enter the values for

the calculations. The data of this key are method

specific to a sample (sample size and identification) or

to a mode (titer, factor, blank, mean value) and the drift

correction applies to all modes. On exit the result is

always recalculated. Values marked with "cond." are

accessible during the conditioning. "**titr." means that

these values can be changed during the titration. They

influence the ongoing determination.


calculation:
Values for result calculations.


statistics:
Calculation of mean values and standard deviation.

The display texts of the Titrino are shown to the left.

The values are the default values.


>calculation

smpl size 1.0 g

**titr.







ident.

**titr.






titer 5 mg/ml

cond.




factor 0.1

cond.

Assignments for the result calculation

Sample size (6-digit number ±X.XXXXX)

Entry using keypad, via balance.

Sample sizes that are inquired automatically in the

titration sequence can be viewed and modified with the

<CALC DATA> key. For calculation the absolute

sample size is used (negative signs that appear, e.g. in

backweighing are ignored).

No inquiry of sample size in the "BLANK" mode.

Sample identification (up to 8 ASCII characters)

Entry using keypad, via balance.

Sample sizes that are inquired automatically in the

titration sequence can be viewed and modified with the

<CALC DATA> key.

No inquiry of sample size in the "BLANK" mode.

Titer(0.0...99.9991 mg/mL)

The titer is written in automatically from the titer

determination modes.

Request only in the "KFT" mode.

Factor (

±

1000000)

in the numerator of the formula. Entries for factors with

different units for results and sample size, see page

33f.


CALC