7 polarizer test – Metrohm 787 KF Titrino User Manual

Page 50

Advertising
background image

3.3. Diagnosis

787 KF Titrino

46

pot.meter dV/dt  10?

• Turn knob ‘dV/dt’ to the right stop and press <Enter>.

Test

sequence:

a) In a first step, the frequency of the RC oscillator (analogue rate) is tested over a period

of 1 second.

b) In a second step, the frequency of the quartz oscillator (digital rate) is tested over a

period of 1 second.

c) If no error is found, after about 3 s it appears

motor-timer test o.k.

• Press <Clear>.

diagnose press key 0...9

7 Polarizer test

• Press <7>.

polarizer test

• Press <Enter>.

dummy resistor 14.3k ?

• Connect resistor switch-box (or suitable resistor 14.3 kΩ 0.1%) using 3.496.5070

cable to ‘Pol’ socket. Switch-box to 14.3 kΩ.

• Press <Enter>.

Test

sequence:

1. An asterisk flashes during the test.
2. In case of an error an error message appears. (If for example the switch-box is not

connected,

error 100

appears).

3. If no error is found, after about 15 s display shows

polarizer test o.k.

• Press <Clear>.

diagnose press key 0...9

• Remove cable and resistor switch-box.