Analysis tool box tools – Bio-Rad Image Lab™ Software User Manual

Page 129

Advertising
background image

Analysis Tool Box Tools

User Guide | 129

The band detection sensitivity options are as follows:

Low Band Detection Sensitivity — sets detection at a low level for images with

prominent bands. Faint bands are not detected with this setting.

High Band Detection Sensitivity — sets detection at a higher level for images that

are faint. Extraneous bands can be removed using the Band Tools in the Analysis

Tool Box. See

Lane and Bands Tools on page 133

.

Custom — enables you to set a value between 1 and 100 to select the best

detection sensitivity for your sample. You can also drag the sliding bar left or right to

set the value.

When Low Band Detection Sensitivity is used, the numerical value is set at 25; when

High Band Detection Sensitivity is used, the value is set at 75.

Molecular Weight Analysis Settings

Molecular Weight Standard — choose any of the many Bio-Rad standards or

another standard you added to your standards list.

Standard Lanes — choose or change the lanes in which the standards are placed.

Regression Method — four regression methods are available. For more

information, see

Regression Calculation Methods on page 253

.

Analysis Tool Box Tools

All Analysis Tool Box tools customize the analyzed data in image files. These tools
are available only when an image file is open. Click a specific image to select among

the many windows that might be open in your workspace.

Note: Some tools delete the existing analysis.

To access a tool

Click any of the toolbox buttons.