5 | Viewing Images

124 | ChemiDoc MP Imaging System with Image Lab Software

Standard Curve

The Standard Curve dialog box displays the best curve fit for the defined standards

and the bands relative to this curve for the lane selected in the image. The tabs at

the bottom of the dialog box display the standard curves for three different

analyses.

Standards appear in green. Unknown bands appear in red. You can toggle the

molecular weight display on the y-axis between linear and log scale by clicking the

Log y-axis box at the upper left. The regression method you chose in Molecular

Weight Analysis Tools appears, as well as the formula (if applicable) and the R

2

value

of the regression method.

Tabs in this window enable you to view the molecular weight standard curve, the

absolute quantity standard curve, or the volume standard curve.