A | Maintenance

214 | ChemiDoc MP Imaging System with Image Lab Software

5.

Place it on a nonabrasive surface so that the glass does not get scratched or
damaged.

Note: Do not put the UV cover directly on the bench. Wear gloves when

touching the lamps.

6.

Rotate the lamp until it becomes loose and the pins come to a vertical position.

7.

Remove the lamp. Install the new lamp by rotating so that the pins are
horizontal and the lamp is tight.

8.

Remove the starter by rotating it counterclockwise, and then pull it out.

9.

Insert a new starter into the holder and rotate clockwise.

10.

.

Reassemble the cover and retighten the screws on both sides.