This chapter includes instructions for maintaining the universal hood in proper

working condition by replacing parts.

UV Transilluminator Lamp and Starter Replacement

Note: The UV filter surface should always be kept clean from the chemical

agents used as gel dyes. Use protective gloves when touching the UV

transilluminator cover.

Depending on usage, the UV bulbs and starters last for many years. Replace bulbs

when you notice them flickering. If a bulb does not turn on when it is new or moved,

replace the bulb starter and test the bulb again.

Three types of bulbs are available. The catalog numbers are listed in

Ordering

Information on page 243

. The standard bulb is 302 nm. Optionally, the 254 nm bulb

is used for cross-linking of protein, and the 365 nm bulb is used to minimize

denaturing of DNA.

To replace the lamps

1.

Turn off the power.

2.

Disconnect the power cord from the universal hood.

3.

Remove the four screws located on the left/right sides of the transilluminator
cover.

4.

Remove the cover with the UV glass by sliding it forward, then lifting up.