6 | Analyzing Images

132 | ChemiDoc MP Imaging System with Image Lab Software

3.

Select the saved crop settings that you want to use and click Load. The red box
resizes and the crop specifications appear on the image.

4.

Right-click and select Crop. The image is cropped to the area specified in the
crop settings you selected.

To delete crop settings

1.

With an image open, click Crop.

2.

Right-click the image inside the red box to open the Crop menu and click
Delete Crop Settings.

3.

Select the crop settings in the dialog box that appears and click Delete.

Inverting Data

Use this button to change the image data from positive to negative. Invert Data is

used for negative stains and zymograms. Intensity values of bands must be greater

than background to perform analysis on the gel. View the gel as a 3-D projection to

determine if the data must be inverted.

Merging Images

Use this button to merge a chemiluminescent blot image with a colorimetric image

of the same blot. If you have used colorimetric prestained standards for a

chemiluminescent blot, you can acquire an epi-white light image of the blot to show

the standards and a chemiluminescent image to show immunodetection. These two

images can then be merged into a combined image with both signals.

Note: Merging images can have an adverse effect on quantification. If accurate

quantification is required, perform analysis on the original, separate images.

Only images of the same size can be merged.